Model Number ARD568370935
Device Problems Corroded (1131); Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.
 
Event Description
On (b)(6) 2021 getinge became aware of an issue with powerled surgical light.As it was stated, there was corrosion formation and paint damage on the light head cover.There was no injury reported, however we decided to report the issue in abundance of caution as any rust or paint particles falling off into sterile field or during procedure may cause contamination.
 
Manufacturer Narrative
Getinge became aware of an issue with powerled surgical light.As it was stated, there was corrosion formation and paint damage on the light head cover.There was no injury reported, however we decided to report the issue in abundance of caution as any rust or paint particles falling off into sterile field or during procedure may cause contamination.It was established that when the event occurred, the surgical light did not meet its specification, since appearance of chipped paint and rust could be considered as technical deficiency, and in this way device contributed to event.We have not received information whether the device was being used for patient treatment at the time when the event occurred.When reviewing similar reportable events for the same device type, we have been able to confirm that the investigated issue has never led to serious injury or worse, to our knowledge.The subject matter experts at the manufacturing site have investigated this type of issue and concluded as follows: all maquet sas products comply with: ¿ iec 60601-1 ed.2.0 & ed.3.0 general requirements for basic safety and essential performance.¿ iec 60601-2-41 particular requirements for the safety of surgical luminaires and luminaire for diagnosis.¿ paint definition pfc066.This procedure defines maquet sas¿s requirements for all painted parts.¿ disinfection products test: the aim of these tests is to detect any incompatibility with disinfectant.The paint chip or paint damages are due to: impacts, collisions (abnormal use).The operating manual includes the instructions to pre-position the arms prior to use, in order to prevent damages.This event is also due to the cleaning product.To prevent any similar incident, it is recommended to avoid the collisions between devices.Visual inspections during the cleaning allow detecting the painting defect, we recommend to perform corrective maintenance to rectify the default after its detection.Minor paint chip can be repaired with touch up paint, nevertheless, the parts impacted by serious damage must be replaced.It is also recommended to avoid excessive friction during cleaning or inappropriate cleaning products.We believe that all remaining devices are performing correctly in the market.Given the circumstances and after our review of complaint ratio behavior of this nature, we shall continue to monitor for any further events of this nature and do not propose any further action at this time.
 
Event Description
Manufacturer's reference number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERLED
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
MDR Report Key11510833
MDR Text Key240644662
Report Number9710055-2021-00101
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberARD568370935
Device Catalogue NumberARD568370935
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/23/2021
Patient Sequence Number1
-
-