MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
|
Back to Search Results |
|
Model Number 97713 |
Device Problems
Intermittent Continuity (1121); Inappropriate/Inadequate Shock/Stimulation (1574); Electro-Static Discharge (2149); Insufficient Information (3190); Patient Device Interaction Problem (4001)
|
Patient Problems
Electric Shock (2554); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant products: product id: 3998, lot#: j0537594v, product type: lead.Other relevant device(s) are: product id: 3998, serial/lot #: (b)(4), ubd: 10-jun-2009, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a patient via manufacturer representative who was implanted with an implantable neurostimulator (ins) for unknown indications for use.It was reported that the scs wires were faulty.Stimulation was alternating between high and low.Patient wanted to discuss surgical repair of the faulty connection.Issue not resolved at this time.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the patient and it was reported that after the back surgery when the patient resumed normal activity per healthcare provider's (hcp)s orders they leaned over at the gorcery stora nd the ins gave them a great shock.Since then it changes voltage with static electricity and also with pressure applied to one spot.No actions were taken to resolve it.
|
|
Manufacturer Narrative
|
Continuation of d10: product id: 3998, lot# j0537594v, product type: lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
Additional information was received.It was reported the circumstances that led to the faulty wires and alternating stimulation was when the patient was implanted with the battery they neglected to connect the wires properly.There were no steps taken to resolve the issue, the patient could push on the connection and make it fluctuate.
|
|
Search Alerts/Recalls
|
|
|