• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD CONNECTA STOPCOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD CONNECTA STOPCOCK Back to Search Results
Catalog Number 394995
Device Problem Filling Problem (1233)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/08/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: a device history record review could not be performed on model 394995 because a lot number was not provided by the customer.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.There are quality controls currently in place to detect this type of defect during the production process.Two quality alerts have been issued for this defect.Further action has not been determined necessary at this time.Our quality team will continue to monitor the manufacturing process for this defect and other emerging trends investigation conclusion: this failure mode is usually presented by the variation of the length of the tube, the actions that have been taken to mitigate the problem, are two generate quality alerts and an inspection for each shift to verify that the lengths of the tube are the indicated ones, in the same way, work continues to find opportunities for improvement.There have been no changes in raw materials or operations that could contribute to this failure mode.We will keep monitoring the manufacturing process and in case any emerging trend is detected, further actions will be taken if necessary.
 
Event Description
It was reported that the bd connecta¿ stopcock experienced kinked tubing.The following information was provided by the initial reporter: the customer has sent us feedback that the experience lately has been that the product kinks below the luer connector much easier than before.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD CONNECTA STOPCOCK
Type of Device
STOPCOCK
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key11516819
MDR Text Key240816580
Report Number9610847-2021-00105
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number394995
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/19/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-