Model Number N/A
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service technician (fse)reported that cardiosave cannot charge battery sometimes.The fse replaced exch pcb, power management, rohs (0670-00-1162 ) and the unit was working and was able to charged the battery.Performed all functional and safety checks to meet factory specifications.The iabp was return to customer and cleared for clinical use.
 
Event Description
It was reported that before use, the cardiosave intra-aortic balloon pump (iabp) does not charge the battery sometimes.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Updated fields: b4, g4, g7, h2, h6 (type of investigation), h10, h11.Corrected field: h4.Analysis of production: (3331) the device history record review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110) the overall 24 month product complaint trend data for the period mar 2019 through feb 2021 was reviewed.There were no triggers identified for the review period.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
MDR Report Key11518521
MDR Text Key240961643
Report Number2249723-2021-00557
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108414
UDI-Public10607567108414
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-0800-55
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/17/2021
Patient Sequence Number1
-
-