DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/23/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service technician (fse)reported that cardiosave cannot charge battery sometimes.
The fse replaced exch pcb, power management, rohs (0670-00-1162 ) and the unit was working and was able to charged the battery.
Performed all functional and safety checks to meet factory specifications.
The iabp was return to customer and cleared for clinical use.
|
|
Event Description
|
It was reported that before use, the cardiosave intra-aortic balloon pump (iabp) does not charge the battery sometimes.
There was no patient involvement, and no adverse event reported.
|
|