The investigation determined that discordant vitros tsh results were obtained from a single patient sample when tested on a vitros 5600 integrated system and a vitros xt7600 integrated system using a vitros tsh reagent lot when compared to non-vitros tsh results.A definitive assignable cause for the discordant vitros tsh results could not be determined with the information provided.Based on historical quality control results, a vitros tsh performance issue is not a likely contributor to the event.There is also no indication of an instrument malfunction and unexpected instrument performance is not a likely contributor to the event.However, an instrument issue cannot be entirely ruled out as a contributing factor as precision testing to assess instrument performance was not performed by the customer at the time of the event.In addition, pre-analytical sample processing could not be ruled out as a contributing factor, as it was established the customer was not following the sample collection device manufactures recommendation for sample centrifugation.It is possible that cellular debris, due to poor sample preparation, was present in the affected sample, although this could not be confirmed.The most likely assignable cause for the discordant vitros tsh results for patient 1 is an unknown sample specific interferent present in the samples drawn from the patient in question.Continual tracking and trending of complaints has not identified any signals that would indicate a potential systematic issue with vitros tsh reagent lot 6400.Email address for contact office is (b)(6).
|