MAUDE Adverse Event Report: SECHRIST INDUSTRIES, INC SECHRIST AIR/OXYGEN GAS BLENDER MODEL 3500CP-G GAS CONTROL UNIT, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SECHRIST INDUSTRIES, INC SECHRIST AIR/OXYGEN GAS BLENDER MODEL 3500CP-G GAS CONTROL UNIT, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 3500CP-G
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/24/2021
Event Type  malfunction  
Manufacturer Narrative
Evaluation of the returned device found the unit is out of calibration. Visual inspection also found one of the two set screws is missing, the fio2 adjustment shaft is topped and not making contact with the double taper needle. The fio2 and balance regulator are out of calibration due to the knob slipping on the shaft. Sechrist recommends completing daily, weekly and semi-annual performance verification on the device to detect any issues and preclude the device from use. In instances where the device is found to function outside of specification, sechrist should be immediately notified for technical assistance. Based on the age of the device, it is unlikely that a manufacturing non-conformity contributed to the reported complaint, and the instructions for use were reviewed and determined to provide adequate instructions and warnings for the safe and effective use of the device. Therefore, no corrective or preventative actions are necessary. All complaints are trended and reviewed by management on a monthly basis. As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted. Manufacturer reference file no. (b)(4).
 
Event Description
Customer reported during clinical use, when they tried to adjust the knob, nothing happened (device did not respond). The mixer was removed from use and replaced with another device. No injury to patient was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSECHRIST AIR/OXYGEN GAS BLENDER MODEL 3500CP-G
Type of DeviceGAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SECHRIST INDUSTRIES, INC
4225 e. la palma ave.
anaheim CA 92807
Manufacturer (Section G)
SECHRIST INDUSTRIES, INC
4225 e. la palma ave.
anaheim CA 92807
Manufacturer Contact
majid mashayekh
4225 e. la palma ave.
anaheim, CA 92807
7145798309
MDR Report Key11537311
MDR Text Key247548292
Report Number2020676-2021-00005
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K023745
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/21/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number3500CP-G
Device Catalogue Number3500CP-G
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/24/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/07/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-