Product event summary: the data files were returned and analyzed.
The files showed at least four applications were performed with a non-returned balloon catheter without any issue on the date of the event.
In conclusion, the clinical issues (arterial puncture, bleed, hematoma) occurred following the procedure.
The case was aborted under general anesthesia.
There is no indication of relation of adverse event to the performance and malfunction of the product.
The physical product was not returned.
If information is provided in the future, a supplemental report will be issued.
|
It was reported that following a successful cryo ablation procedure, when the sheath was removed from the patient, oozing was noted.
Groin pressure was maintained.
The patient became unstable.
A cardiac ultrasound was performed to exclude tamponade.
Ultrasound of the groin showed pulsating blood flow into a hematoma.
A femoral compression system was applied.
It was concluded an arterial puncture or rupture had occurred.
The patient was transferred to surgery for groin exploration and vascular closure and repair.
No further patient complications have been reported as a result of this event.
|