WILLIAM COOK EUROPE GUNTHER TULIP NAVALIGN JUGULAR VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-1-JUG-TULIP
Device Problems Structural Problem (2506); Appropriate Term/Code Not Available (3191)
Patient Problems Internal Organ Perforation (1987); Pain (1994); Perforation of Vessels (2135); Loss of Vision (2139); Numbness (2415); Cognitive Changes (2551); Insufficient Information (4580)
Event Type  Injury  
Manufacturer Narrative
Non-healthcare professional.Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog number and lot number are unknown, however, the alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.
 
Event Description
It is alleged that the patient received a gunther tulip on (b)(6) 2012.It is alleged that the patient was injured without further explanation.Hospital and medical records have been requested, but not yet provided.
 
Manufacturer Narrative
G4: 510(k) - k172557.Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: vena cava/organ perforation, numbness, impaired cognition/vision, pain.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Unknown if the reported numbness, impaired cognition/vision, pain are directly related to the filter and unable to identify a corresponding failure mode at this point in time.20 devices in lot.No relevant notes on work order.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Patient allegedly received an implant on (b)(6) 2012 via the right internal jugular vein due to deep vein thrombosis (dvt) and pulmonary embolism (pe).Patient is alleging vertebral body perforation.The patient further alleges "numbness of extremities, impaired cognitive function, vision impairment/ loss, and back pain." report from ct (computed tomography): "there is a greenfield type ivc filter in place within the mid ivc.The superior most aspect of the filter is below the right and left renal veins.There is a lumbar spine scoliosis convexity towards the right which deviates the ivc.The ivc filter is normally oriented within the lumen of the ivc.The four most distal anchoring prongs projecting outside the confines of the ivc wall at the 12:00, 3:00, 6:00 and 9:00 positions.The 12:00 anchoring prongs appears to extend up against the wall of the third portion of the duodenum.The three other anchoring prongs projecting into the retroperitoneal fat.".
 
Manufacturer Narrative
Investigation: investigation is reopened due to additional information provided.Per quality engineering review, the additional information provided for this complaint does not change the previous investigation conclusion.Therefore, no new investigation activities will be conducted at this time.Cook will reopen the investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.(b)(4) devices in lot.The associated work order was reviewed.No related/relevant notes were documented.One other complaint on lot for different issue.Product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.  a follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP NAVALIGN JUGULAR VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
henriette s. christiansen
sandet 6, dk-4632
bjaeverskov 
56868686
MDR Report Key11547694
MDR Text Key245610416
Report Number3002808486-2021-00993
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/21/2014
Device Catalogue NumberIGTCFS-65-1-JUG-TULIP
Device Lot NumberE2849666
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient SexFemale
-
-