MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 32MM X 48MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 32MM X 48MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Model Number 71337648
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/26/2021
Event Type  malfunction  
Event Description
It was reported that during a hip surgery, the insert couldn't be implanted after several times.Finally a backup ceramic insert was used to complete the surgery.No surgical delay or injury to the patient was reported.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was returned for evaluation.A visual inspection was completed but could not confirm the state failure.There are gouges and minor damage to the liner.The liner exhibits deep scratches/deformation on the id and id rim.A dimensional analysis could not be completed due to damage to the liner from attempting to insert it into the shell.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 20 DEG XLPE ACET LNR 32MM X 48MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11548609
MDR Text Key241675008
Report Number1020279-2021-02304
Device Sequence Number1
Product Code MBL
UDI-Device Identifier03596010598479
UDI-Public03596010598479
Combination Product (y/n)N
PMA/PMN Number
K113848
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71337648
Device Catalogue Number71337648
Device Lot Number19CM17078
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-