• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL LTV 1150 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL LTV 1150 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number LTV 1150
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 01/12/2021
Event Type  Death  
Manufacturer Narrative
The equipment was received in vyaire facility for evaluation.The unit is placed on extended test/run-in.No root cause has been determined at this time, since the investigation is still ongoing.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
It was reported to vyaire that a patient passed away while connected to the lap top ventilator 1150 and requested to check the event trace.At this time, no information is available regarding whether or not the device had anything to do with the patient's passing.
 
Manufacturer Narrative
Device evaluation: the vyaire failure analysis laboratory received the suspect component and performed a failure investigation.The service technician was able to verify the reported problem "missing multiple screws".Visual inspection reveals three missing screws on the upper protective boot.The unit passed extended tests at run-in settings.The tidal volume & flow performance failed for non-conformance with measured bias flow.The measured bias flow was 11.10 while the expected is 11.00.V-home was performed.The rear weldment was replaced and all screws were installed at ltv (lap top ventilator) completion.The unit passed all testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LTV 1150 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa IL 60045
MDR Report Key11554874
MDR Text Key241789453
Report Number2021710-2021-13572
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00845873002726
UDI-Public(01)00845873002726(11)20151231
Combination Product (y/n)N
PMA/PMN Number
K060647
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberLTV 1150
Device Catalogue Number18984-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2021
Date Manufacturer Received04/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Death;
-
-