Vyaire medical file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
Device evaluation: g3, g6, h2, h3, h6 and h10.Result of investigation: the exact root cause is not determinable.Most likely the issue is caused by a hardware issue on the epc.Despite many tests performed by imt ag on similar returned parts, the issue was still not reproducible.Investigation will be continued under (b)(4).As the failure rate is within the expected range as per our risk files, decided to close this complaint.
|