PHYSIO-CONTROL, INC. - 3015876 LIFEPAK 15 DEFIBRILLATOR/MONITOR AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/03/2021
Event Type  malfunction  
Manufacturer Narrative
Physio-control contacted the customer to request additional information on the patient. No response has been received from the customer. Patient fields in which information was not provided were intentionally left blank. Physio-control evaluated the customer's device and was unable to duplicate the reported issue. After observing proper device operation through functional and performance testing, the device was returned to the customer for use. The cause of the reported issue could not be determined.
 
Event Description
The customer contacted physio-control to report that paddles lead ecg was not visible on their device during a patient event. As a result, defibrillation therapy could have been delayed or unavailable, if needed. There were no reports of any adverse effects to the patient as a result of the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameLIFEPAK 15 DEFIBRILLATOR/MONITOR
Type of DeviceAUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key11558192
MDR Text Key242288450
Report Number0003015876-2021-00631
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation
Type of Report Initial
Report Date 03/23/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number15
Device Catalogue Number99577-001957
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2021
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/23/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/07/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-