Model Number 88-9199
Device Problem Fire (1245)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/17/2021
Event Type  malfunction  
Event Description
During surgery, crna noted a flame coming from the robot arm where the robotic instrument connects to the robotic arm.Surgery immediately stopped unit flame was put out and affected surgical instruments were removed from field.New monopolar cord and monopolar scissors were added back to the field.Flame and burn on drape was not near the patient or her skin.Skin assessed after drape was removed and no injury noted to patient.Md notified.Spoke with staff involved.(b)(6) states saw flame and notified room of fire.Davinci instrument removed from robot arm 1 by dr.(b)(6) and handed to (b)(6).(b)(6) able to blow out flame.States it was the size of a large birthday candle.Biomed notified and check valley lab generator.Intuitive notified and will send field rep tomorrow to check robot.Csp notified of complete separation in bovie cord.Bovie cord from cardinal care fusion v.Muller , item 88-9199 cord with noted number (b)(4) as well.Cardinal care fusion has not been notified of incident.I do not have a rep or contact name, these items are ordered by csp and placed on our robotic trays.Photos on file.Instruments in possession of (b)(6) at this time.Will be provided to manufacturer upon request with proper chain of custody documentation.Fda safety report id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V. MUELLER BOVIE CORD
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
CARDINAL/CAREFUSION 2200, INC.
MDR Report Key11562249
MDR Text Key242038793
Report NumberMW5100219
Device Sequence Number2
Product Code GEI
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 03/22/2021
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Date FDA Received03/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number88-9199
Device Catalogue Number88-9199
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age37 YR
Patient Weight112
-
-