• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP VENTRICULAR (ASSIST) BYPASS Back to Search Results
Device Problem Malposition of Device (2616)
Patient Problem Heart Failure/Congestive Heart Failure (4446)
Event Date 01/01/2020
Event Type  Injury  
Manufacturer Narrative
Product event summary: a pump and an outflow graft with unknown serial numbers were not returned for evaluation. Review of the controller log files was not performed since log files covering the reported event date were not available for analysis. There is no evidence to suggest that a device malfunction caused or contributed to the reported event. Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patient's complex post-operative course. There are possible patient, pharmacological and procedural factors that may have contributed to this event. Additional products: heartware ventricular assist system outflow graft, model #: unk / catalog #: unk / expiration date: unk / serial or lot#: unk, udi #: asku, mfg date: unk, (b)(4). This event was reported in the q4 2020 intermacs data registry that tracks clinical outcomes of patients on ventricular assist device (vad) support. The data registry does not contain device identifying information or event date and therefore cannot be correlated to any previously received report of the event. Based on the provided data, device analysis will not be possible and no further information will likely be made available concerning the event. Investigation of this event is completed and the file will be closed. If new information is received, the file will be re-opened and a supplemental will be submitted. If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient was hospitalized due to worsening right heart failure symptoms. A computed tomography angiography (cta) confirmed narrowing or compression of the outflow graft in addition to suboptimal ventricular assist device (vad) inflow cannula position. The patient underwent an unspecified surgical procedure. The vad and outflow graft remain in use. No further patient complications have been reported as a result of this event. This event was reported in the q4 2020 intermacs data registry that tracks clinical outcomes of patients on ventricular assist device (vad) support. The data registry does not contain device identifying information or event date and therefore cannot be correlated to any previously received report of the event. Based on the provided data, device analysis will not be possible and no further information will likely be made available concerning the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP
Type of DeviceVENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key11565497
MDR Text Key257856428
Report Number3007042319-2021-02863
Device Sequence Number1
Product Code DSQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/23/2021
Was Device Evaluated by Manufacturer? No Answer Provided
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 03/24/2021 Patient Sequence Number: 1
-
-