Model Number 7015
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2020
Event Type  malfunction  
Manufacturer Narrative
Initial reporter facility name: hospital affiliated to school of medicine, (b)(6).Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were kinks 12cm and 14.9cm distal from the strain relief.The balloon was loosely folded.Microscopic inspection revealed tip damage.There was a longitudinal tear 1mm long starting 2mm proximal from the proximal marker band.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event Description
Reportable based on device analysis completed on (b)(6) 2021.It was reported that shaft kink occurred.The target lesion was located in the left circumflex artery.A 3.0mm x 15mm quantum maverick balloon catheter was advanced for dilation.However, during procedure, it was noted that the balloon delivery shaft was kinked.The procedure was completed with another of the same device.There were no patient complications and the patient was stable.However, returned device analysis revealed balloon longitudinal tear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key11565669
MDR Text Key244444525
Report Number2134265-2021-03856
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08714729392781
UDI-Public08714729392781
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P860019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 03/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/12/2023
Device Model Number7015
Device Catalogue Number7015
Device Lot Number0025346537
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/04/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age52 YR
Patient Weight78
-
-