MAUDE Adverse Event Report: ORTHO-CLINICAL DIAGNOSTICS VITROS CHEMISTRY PRODUCTS PHYT SLIDES; IN VITRO DIAGNOSTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ORTHO-CLINICAL DIAGNOSTICS VITROS CHEMISTRY PRODUCTS PHYT SLIDES; IN VITRO DIAGNOSTICS Back to Search Results
Catalog Number 8298671
Device Problem No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/28/2021
Event Type  malfunction  
Event Description
A customer obtained higher than expected vitros phenytoin (phyt) results from vitros therapeutic drug monitoring performance verifiers (tdm pv) processed using vitros chemistry products phyt slides on two vitros 4600 chemistry systems.Analyzer 46001169: tdmpvi lot r7687 results of 12.1, 10.3, 11.1, 10.5, 10.5, 10.9, 11.0, 11.2 and 11.7 ug/ml vs the expected result of 8.3 ug/ml.Tdm pvii lot t7688 results of 18.0, 17.6, 16.9, 17.3, 17.5, 18.0, 25.1, 19.4, 19.1, 18.2, 17.8, 17.1, 17.7, 17.4, 16.9, 16.8, 16.8, 17.3, 22.1, 23.2, 27.1, 23.4, 22.1, 17.9, 28.6, 25.3, 19.2, 19.9, 20.2, 17.0, 17.3 and 17.4 ug/ml vs the expected result of 14.0 ug/ml.Analyzer 46000903: tdm pvii lot t7688 result of 17.3 ug/ml vs the expected result of 14.0 ug/ml.Biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected on patient samples.The customer, through the distributor field engineer, indicated that patient results were reported outside of the laboratory while running vitros phyt during the timeframe of the event.However, the results were not questioned by a physician and no repeats were requested for any of the samples.It cannot be confirmed that patient results would not be affected if the issue were to recur undetected.There is no allegation of patient harm as a result of this event.This report is number 40 of 42 mdr¿s for this event.Forty-two (42) 3500a forms are being submitted for this event as 42 devices were involved.This report corresponds to ortho clinical diagnostics inc.Complaint (b)(4).
 
Manufacturer Narrative
The investigation determined that higher than expected vitros phenytoin (phyt) results were obtained from vitros therapeutic drug monitoring performance verifiers (tdm pv) processed using vitros chemistry products phyt slides lot 2621-0175-3482 on two vitros 4600 chemistry systems.The assignable cause of the event is unable to be determined with the information provided.Historical quality control results show inaccuracy for vitros tdm pvii, and imprecision for both levels of vitros tdm pv fluids.This indicates that vitros phyt lot 2621-0175-3482 is not performing as expected in combination with the vitros 4600 chemistry systems.The customer did not process any of the requested diagnostic precision tests.Therefore, it is unknown if a regent issue or an analyzer issue contributed to the higher than expected results.Continual tracking and trending does not indicate a systemic issue with vitros phyt lot 2621-0175-3482.The fluid and reagent handling protocol cannot be ruled out as a contributor to the event.The customer processed multiple results using onboard expired reagents.Additionally, it is unknown if the fluid or reagent warm up times used were acceptable per the vitros phyt instructions for use.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITROS CHEMISTRY PRODUCTS PHYT SLIDES
Type of Device
IN VITRO DIAGNOSTICS
Manufacturer (Section D)
ORTHO-CLINICAL DIAGNOSTICS
100 indigo creek drive
rochester NY 14626
MDR Report Key11567826
MDR Text Key280508204
Report Number1319809-2021-00066
Device Sequence Number1
Product Code DIP
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial
Report Date 03/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/15/2021
Device Catalogue Number8298671
Device Lot Number2621-0175-3482
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-