The user was wearing a lab coat and gloves at the time of the incident.The sysmex xn-3000 instructions for use (ifu), chapter 2 - safety information, section 2.4 - avoiding infections, states, "when performing any task on the instrument, such as testing, maintenance, preparation, or post processing, be sure to wear adequate personal protective equipment, such as protective gloves, a protective mask, protective eyewear, and a lab coat." the user was not wearing safety glasses.Section 2.4 goes on further stating users are to connect waste tubing to a waste fluid tank or other dedicated waste receptacle.To avoid the risk of waste spillage, users are advised to secure waste tubing in place and regularly verify waste tubing remains properly secured.Waste tubing and drain maintenance is the responsibility of the laboratory.Waste tubing was not properly secured in the drain.A sysmex service engineer (se) was dispatched.The se inserted all waste tubing directly into the floor drain and secured all tubing to prevent recurrence.No systemic product deficiency was identified.
|
While investigating an analyzer leak during operation, the user observed waste tubing was not securely inserted into the drain causing waste to splash into the user's eyes.The user's eyes were immediately flushed by the laboratory's eyewash station, and the user sought medical attention from an exposure nurse.The user was placed on the (b)(6) protocol, which included testing and the administration of the prophylactics isentress and truvada.The user suffered eye irritation from the waste splash and/or the eyewash, as well as nausea, vertigo, and vomiting from the prophylactics.It is unknown if the user sought medical care for the eye irritation.
|