BECTON DICKINSON, S.A. BD PLASTIPAK CONCENTRIC LUER LOCK SYRINGE PISTON SYRINGE Back to Search Results
Catalog Number 300865
Device Problem Inappropriate Audible Prompt/Feedback (2280)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2020
Event Type  malfunction  
Manufacturer Narrative
Medical device expiration date: unknown. Device manufacture date: unknown. Investigation summary: no photos or physical samples that display the reported condition were available for investigation. A device history review could not be completed as no batch number was provided. Based on the available information we are not able to identify a root cause at this time. Complaints received for this device and reported condition will continue to be tracked and trended. Our quality team regularly reviews the collected data for identification of emerging trends. Based on no sample, the investigation concluded, bd was not able to verify the indicated failure.
 
Event Description
It was reported that the syringe pump alarm went off during use with the bd plastipak¿ concentric luer lock syringe, despite there being no occlusion. The following information was provided by the initial reporter: "infusion pump alarming "occlusion" with no obvious source of "occlusion". Replacement syringe prepared and no issue with occlusion using fresh syringe. On examination of the original bd platipak syringe it appeared to be sticky. No injury or illness".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameBD PLASTIPAK CONCENTRIC LUER LOCK SYRINGE
Type of DevicePISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer (Section G)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key11579701
MDR Text Key242756723
Report Number3003152976-2021-00171
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/23/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number300865
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/02/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-