• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HOL ML 5MM ENDO APPLIE; APPLIER, SURGICAL, CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HOL ML 5MM ENDO APPLIE; APPLIER, SURGICAL, CLIP Back to Search Results
Model Number IPN915189
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/05/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that the pivot pin was found detached before use.Therefore, a new applier was used instead.
 
Manufacturer Narrative
Qn# (b)(4).The dhr for the returned instrument was reviewed and found completely without any irregularities.This instrument was produced at the tecomet, inc.Kenosha, wi facility as part of a (b)(4) pc.Lot in april of 2020.Evaluation of this instrument shows that the jaw pivot pin is slightly sticking out of the outer tube assembly.Further evaluation shows that the jaws are misaligned in the closed position.We are able to validate this complaint.After the initial evaluation this instrument was dis-assembled in order to evaluate its internal components and it was found that the drive rod (n00185) fingers are also damaged.It is suspected that the damaged drive rod fingers are what caused the jaws to become misaligned and for the jaw pivot pin to be slightly sticking out of one side of the outer tube assembly.Mishandling of this device at the end user's facility is suspected.All (b)(4) instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.
 
Event Description
It was reported that the pivot pin was found detached before use.Therefore, a new applier was used instead.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOL ML 5MM ENDO APPLIE
Type of Device
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key11581182
MDR Text Key243370941
Report Number3011137372-2021-00065
Device Sequence Number1
Product Code GDO
UDI-Device Identifier24026704696813
UDI-Public24026704696813
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 03/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN915189
Device Catalogue Number544965
Device Lot Number06B2098678
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2021
Date Manufacturer Received04/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-