Model Number IPN915189 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
|
Event Description
|
It was reported that the pivot pin was found detached before use.Therefore, a new applier was used instead.
|
|
Manufacturer Narrative
|
Qn# (b)(4).The dhr for the returned instrument was reviewed and found completely without any irregularities.This instrument was produced at the tecomet, inc.Kenosha, wi facility as part of a (b)(4) pc.Lot in april of 2020.Evaluation of this instrument shows that the jaw pivot pin is slightly sticking out of the outer tube assembly.Further evaluation shows that the jaws are misaligned in the closed position.We are able to validate this complaint.After the initial evaluation this instrument was dis-assembled in order to evaluate its internal components and it was found that the drive rod (n00185) fingers are also damaged.It is suspected that the damaged drive rod fingers are what caused the jaws to become misaligned and for the jaw pivot pin to be slightly sticking out of one side of the outer tube assembly.Mishandling of this device at the end user's facility is suspected.All (b)(4) instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.
|
|
Event Description
|
It was reported that the pivot pin was found detached before use.Therefore, a new applier was used instead.
|
|