Concomitant medical products: product id 8781, serial# (b)(4), product type catheter, product id 8780, serial# (b)(4), implanted: (b)(6) 2021, product type catheter.Other relevant device(s) are: product id: 8781, serial/lot #: (b)(4), ubd: 09-jul-2021, udi#: (b)(4); product id: 8781, serial/lot #: hg3h0y008, ubd: 12-jun-2021, udi#: (b)(4).The catheters were returned and analysis identified an anomaly to the catheter/guide wire.If information is provided in the future, a supplemental report will be issued.
|
Information was received from a healthcare provider (hcp) via a company representative regarding a patient receiving gablofen (500 mcg/ml at 75 mg/day) via an implanted pump.It was reported that during the initial implant procedure the doctor was advancing the catheter when the wiring was coming out and the catheter bent.He was unable to get placement, so asked for another catheter to be opened.The same issue occurred, and he asked for another catheter to be opened.He was then able to get placement and the wiring didn¿t come out.The environmental, external, or patient factor that may have led or contributed to the issue was noted to be ¿patient¿s spinal anatomy¿.The issue was resolved, and it was indicated that the hcp had no further information to provide regarding the event.
|