• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72200616
Device Problems Material Deformation (2976); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Event Description
It was reported that, during use, the "mdu powermax elite shaver" had a damaged cable and the motor stalled and did not come out.Although the surgery was not delayed, it is unknown if a back-up device was available to complete the procedure.No patient injuries or other complications were reported.No patient injuries or other complications were reported.No further details are available.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A relationship, if any, between the subject device and the reported event could not be determined.No containment or corrective actions are recommended at this time.A complaint history review concluded this was a repeat issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11585844
MDR Text Key243815317
Report Number1643264-2021-01196
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010555472
UDI-Public03596010555472
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 05/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200616
Device Catalogue Number72200616
Was Device Available for Evaluation? Yes
Date Manufacturer Received05/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-