• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 150 -EUR4

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 150 -EUR4 Back to Search Results
Model Number 27063
Device Problems Device Displays Incorrect Message (2591); Battery Problem (2885)
Patient Problem Insufficient Information (4580)
Event Date 01/04/2021
Event Type  malfunction  
Manufacturer Narrative
The astral device was returned to resmed for an investigation.Review of the device data logs confirmed the reported complaint and revealed an error message related to a battery charger fault (sf180) and a battery inoperable alarm.Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the sf180 and internal battery degraded warning alarm were due to an isolated component failure within the device battery assembly while the internal battery inoperable alarm was due to an intermittent connection with the battery.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed an internal battery degraded warning alarm.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 -EUR4
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key11586590
MDR Text Key243609916
Report Number3007573469-2021-00455
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00619498270637
UDI-Public(01)00619498270637(10)1352295
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27063
Device Catalogue Number27063
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/29/2021
Distributor Facility Aware Date03/09/2021
Device Age25 MO
Date Report to Manufacturer03/29/2021
Date Manufacturer Received03/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-