DATASCOPE CORP. - FAIRFIELD SENSATION PLUS 8FR. 50CC IAB SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0684-00-0576-01 |
Device Problem
Difficult to Advance (2920)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Complete initial reporter name: (b)(6), rn.
The device was not returned and could not be evaluated.
It was discarded by the user.
We are unable to confirm the reported event.
If new information becomes available, a supplemental report will be submitted.
Complaint record id # (b)(4).
|
|
Event Description
|
It was reported that during insertion of the intra-aortic balloon (iab), the customer could not advance the iab any further.
The console was started and it was noticed that only half of the iab was inflating.
The iab was removed and replaced with a new one which worked normally without issue.
There was no patient harm or adverse event reported.
|
|
Search Alerts/Recalls
|
|
|