Model Number CV-180
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/05/2021
Event Type  malfunction  
Manufacturer Narrative
During a call with tac (technical assistance center) support engineer, customer conveyed that the integration system boom (route signal) that is being used with the cv-180 is a non-olympus unit. In addition, the cv- 180 was acquired through a third party. Customer was advised to contact their boom provider, manufacturer for the boom system. The cv-180 was not returned for evaluation. To date, there are no other issues reported. This report will be supplemented accordingly following investigations.
 
Event Description
Short, intermittent image was reported on the device. The red cable on the integration system is no longer working, when being moved, it can manipulate the image. There was no patient involvement, no user injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameEVIS EXERA II VIDEO SYSTEM CENTER
Type of DeviceVIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11592504
MDR Text Key245893834
Report Number8010047-2021-04366
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation
Type of Report Initial,Followup
Report Date 04/28/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberCV-180
Device Catalogue NumberCV-180
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received04/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/22/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-