• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 ADAPTER SLEEVES 12/14 +8; FEMORAL SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 ADAPTER SLEEVES 12/14 +8; FEMORAL SLEEVE Back to Search Results
Catalog Number 999800318
Device Problems Degraded (1153); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Foreign Body Reaction (1868); Osteolysis (2377); Muscle/Tendon Damage (4532); Unspecified Tissue Injury (4559); Swelling/ Edema (4577); Insufficient Information (4580)
Event Date 12/01/2020
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.This hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Asr litigation records received.Records alleges injury, economic loss, loss of services and loss of consortium.Doi: (b)(6) 2007 dor: (b)(6) 2020 left hip.
 
Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.This hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: previous investigations that have included manufacturing record evaluations (mre) since the asr platform was launched have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.
 
Event Description
Failed left total hip replacement secondary to adverse local tissue reaction, associated with metal-on-metal large head total hip replacement.There was a relatively large contained lytic lesion in the inferior aspect of the acetabulum.Operative notes reported that there was a significant taper corrosion on the taper which extended beyond the neck of the femoral component, that is, the corrosion extended outside of the taper.A lytic lesion which had been debrided.Operative findings indicated that there was a moderate amount of fluid in the hip joint.There was also a pseudotumor component which was solid.There was a large contained lytic defect in the inferior half of the acetabulum.The soft tissues were in poor condition.The gluteus medius tendon looked good, but the minimus tendon was actually degenerative and partially avulsed.Doi: (b)(6) 2007.Dor: (b)(6) 2020.Left hip.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Udi (b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADAPTER SLEEVES 12/14 +8
Type of Device
FEMORAL SLEEVE
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key11598888
MDR Text Key243342792
Report Number1818910-2021-06676
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
PMA/PMN Number
K040627
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 03/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2010
Device Catalogue Number999800318
Device Lot Number1972609
Was Device Available for Evaluation? No
Date Manufacturer Received07/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Treatment
ASR ACETABULAR CUPS 60.; ASR UNI FEMORAL IMPL SIZE 53.; SUMMIT POR TAPER SZ10 STD OFF.; UNK HIP ACETABULAR CUP ASR.; UNK HIP FEMORAL HEAD METAL ASR.; UNK HIP FEMORAL SLEEVE ASR.; UNKNOWN HIP FEMORAL STEM.
Patient Outcome(s) Required Intervention;
Patient Age63 YR
-
-