The valve was returned for evaluation: device history record- lot crkcmy, conformed to the specifications when released to stock failure analysis: the valve was visually inspected; no defects were noted.The position of the cam when valve was received was 70mmh2o.The valve passed the test for programming, flushing, leaking, reflux and pressure.The catheter was irrigated, no occlusions noted.No root cause could be determined as the technician could not confirm any problem with the valve at the time of investigation.The possible root cause for the issue reported by the customer, could be due to biological debris and protein build up interfering with the valve mechanism.No occlusion was noted during the investigation of the valve.
|