Model Number BELLAVISTA 1000
Device Problem No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2021
Event Type  malfunction  
Manufacturer Narrative
Vyaire medical file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.However, by looking at the log file and o2 gas path test screenshots provider by the customer the "flow o2" reading in the test was not 0, the o2 dosing valve is leaking.No root cause has been determined at this time, since the investigation is still ongoing.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported that the bellavista 1000 alarmed 379 "no o2 dosing possible".The customer confirmed that there was no patient involvement from the reported event.
 
Manufacturer Narrative
Result of investigation: it was determined that the root cause of the issue was the leaking o2 dosing proportional valve.O2 path self test is failing due to leaking o2 dosing proportional valve.As a resolution o2 dosing valve needs to be replaced.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BELLAVISTA
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
IMTMEDICAL AG
gewerbestrasse 8
buchs, 9470
SZ  9470
MDR Report Key11599984
MDR Text Key244184186
Report Number3004553423-2021-00926
Device Sequence Number1
Product Code CBK
UDI-Device Identifier07640149380019
UDI-Public(01)07640149380019
Combination Product (y/n)N
PMA/PMN Number
K163127
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 03/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBELLAVISTA 1000
Device Catalogue Number301.100.000
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-