Device Problem Break (1069)
Patient Problems Fistula (1862); Perforation of Vessels (2135)
Event Date 12/23/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received a report that the pushwire soldered to the proximal marker of the device fractured. The deployment wire was retained in the patient, and during an attempted retrieval of the wire, the proximal sharp wire fragment perforated the cavernous internal carotid artery (ica). This led to a direct carotid-cavernous fistula. Ultimately, the wire fragment was retrieved using a loop snare by snaring the distal end of the wire. Facility surgery was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePIPELINE FLEX EMBOLIZATION DEVICE
Type of DeviceINTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
david gustafson
9775 toledo way
irvine, CA 92618
7635149628
MDR Report Key11603450
MDR Text Key243526939
Report Number2029214-2021-00383
Device Sequence Number1
Product Code OUT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100018.S011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation
Type of Report Initial,Followup
Report Date 04/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/23/2021
Was Device Evaluated by Manufacturer? No Answer Provided
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 04/01/2021 Patient Sequence Number: 1
-
-