Model Number 71751129
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/12/2021
Event Type  Injury  
Event Description
It was reported that during surgery a ho loan kit was used and several inconvenient were noticed.The sureshot humeral drill guide probe failed.The proximal jig was faulty causing all of the proximal screws to miss the nail.All screws ended up being done by hand making perfect circles with both distal screws and proximal screws.It was reported a significant the case ended up delayed for over 2 hours.The procedure was completed with the same device and no injuries were reported.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.The clinical/medical evaluation concluded that per the complaint details, we are currently unable to rule out a procedural variance as a contributing factor to the reported event which does not represent a device malfunction.Based on the information provided, all screws were done by hand making perfect circles with both distal screws and proximal screws.According to the report, the procedure was completed with the same device with a two-hour delay.Since there were no injuries reported, no further clinical/medical assessment is warranted at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUMERAL NAIL DRILL GUIDE
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11616781
MDR Text Key243891523
Report Number1020279-2021-02693
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010515414
UDI-Public03596010515414
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71751129
Device Catalogue Number71751129
Was Device Available for Evaluation? No
Date Manufacturer Received07/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other; Required Intervention;
-
-