• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMTMEDICAL AG BELLAVISTA; VENTILATOR, CONTINOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMTMEDICAL AG BELLAVISTA; VENTILATOR, CONTINOUS, FACILITY USE Back to Search Results
Model Number BELLAVISTA 1000 US
Device Problem No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2021
Event Type  malfunction  
Manufacturer Narrative
The log file shows multiple alarm 388 (no o2 dosing possible) with 271 (o2 supply fail - no o2 dosing possible) but no 338.It was then advised to replace the inspiration block.At this time, the suspected device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported that the bellavista 1000 us was showing alarm 388 - "no o2 dosing possible" during use on a patient.There was no patient harm or injury as reported.
 
Manufacturer Narrative
Device evaluation: results of investigation: analysis on the log file shows that the issue was due to a defective o2 pressure regulator.Most likely restriction of the p2 regulator hole caused by injection molding deburrs during manufacturing.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BELLAVISTA
Type of Device
VENTILATOR, CONTINOUS, FACILITY USE
Manufacturer (Section D)
IMTMEDICAL AG
gewerbestrasse 8
buchs, 9470
SZ  9470
MDR Report Key11616878
MDR Text Key243940941
Report Number3004553423-2021-00941
Device Sequence Number1
Product Code CBK
UDI-Device Identifier07640149388183
UDI-Public(01)07640149388183(11)20200827
Combination Product (y/n)N
PMA/PMN Number
K163127
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 03/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBELLAVISTA 1000 US
Device Catalogue Number301.100.030
Was Device Available for Evaluation? Yes
Date Manufacturer Received05/31/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-