Model Number BELLAVISTA 1000 US |
Device Problem
No Flow (2991)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/09/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The log file shows multiple alarm 388 (no o2 dosing possible) with 271 (o2 supply fail - no o2 dosing possible) but no 338.It was then advised to replace the inspiration block.At this time, the suspected device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
The customer reported that the bellavista 1000 us was showing alarm 388 - "no o2 dosing possible" during use on a patient.There was no patient harm or injury as reported.
|
|
Manufacturer Narrative
|
Device evaluation: results of investigation: analysis on the log file shows that the issue was due to a defective o2 pressure regulator.Most likely restriction of the p2 regulator hole caused by injection molding deburrs during manufacturing.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|