DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE E/F PLG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-55 |
Device Problem
Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/09/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue.The fse ordered a replacement reel for the defective power supply cable, and upon receipt, repairs can continue.Additionally, the fse performed a successfully pm maintenance service which will require to be performed again following completion of repairs of the iabp unit.Additional information is being requested with regard to the repair and status of the iabp unit.A supplemental report will be submitted when this information is provided to us.The initial reporter named is a getinge employee who has different contact details from that of the event site, and has the following contact information: (b)(6).
|
|
Event Description
|
It was reported that during a preventative maintenance (pm) service repair performed by a getinge field service engineer (fse), the cardiosave intra-aortic balloon pump (iabp) had a defective power supply cable.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device (10): a getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue.The fse replaced the line cord with type e/f plug.The fse performed functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.Customer contact information was provided and is as follows: (b)(6), biomedical engineer, (b)(6).
|
|
Search Alerts/Recalls
|
|
|