Model Number BELLAVISTA 1000 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Vyaire medical file identification: (b)(4).Device evaluated by mfr: 81 other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
The customer reported that the bellavista 1000 has technical failure 401 "inspiration valve or device leaky".There was no information of patient involvement reported from this event.
|
|
Manufacturer Narrative
|
Result of investigation: no failure detected.Most likely a leak on o2 cell port causes tf 401 which was later arrested before service visit.Machine tested by vyaire field service engineer.As per wo-00016083 "checked unit.Also found that the unit software was at 6.0.1100.0.Upgraded instrument.Tested unit.Unit working to specification.
|
|