BAXTER HEALTHCARE CORPORATION EVACUATED CONTAINER; APPARATUS, SUCTION, SINGLE PATIENT USE, PORTABLE, NONPOWERED
|
Back to Search Results |
|
Catalog Number 1A8504 |
Device Problem
Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that eight (8) evacuated containers, 1000 ml were cracked.This was observed prior to use.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
One (1) device was received for evaluation.The other seven (7) samples were not received and therefore, could not be evaluated.A visual inspection was performed which observed a collision bruise and blister.The reported condition was verified.The cause of the condition was due to a manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|