Turnpike was returned to vsi/teleflex for evaluation.Lot: 704744 was reviewed.There are no non-conformances related to this lot therefore supporting the device met material, assembly and performance specifications.Additional information was requested form the account.No response was received.Only an image of the turnpike was sent confirming damages to the tip.The tip was unable to be snared out of the vessel.A stent was placed in the left main to keep the piece of the tip in place against the wall of the artery.No blood particulates were noted on the shaft.The tip of the catheter was severely damaged, torqued, and fatigued.Approximately 0.2 cm of tip material was observed to be sheared off from the distal section of the tip.The shaft of the catheter had a minor kink.Length, o.D., o.D at the distal section, were measured and catheter met specification per drawing.No other damages were noted.
|