Model Number R SERIES
Device Problem Unable to Obtain Readings (1516)
Patient Problem Necrosis (1971)
Event Date 04/01/2021
Event Type  Death  
Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that while attempting to defibrillate a patient (age & gender unknown), the device was unable to obtain an ecg signal.Complainant indicated that the patient subsequently expired.
 
Manufacturer Narrative
Zoll medical singapore evaluated the device and the customer's report was not replicated or confirmed.The device was put through extensive testing including full functional testing and ecg stress testing via leads and pads testing without duplicating the report.The activity log indicated that the device recognizes cpr feedback and a valid patient impedance, which would have displayed a pad signal if the device was set to the correct ecg view.Unfortunately, the activity log was unable to confirm the default lead view or what lead view that was selected at the time.The device was recertified and returned to the customer.Review of the device activity logs did not show any faults that could be associated with customer report.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key11646532
MDR Text Key244804264
Report Number1220908-2021-01212
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
PMA/PMN Number
K060559/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 04/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Death;
-
-