It was reported during a knee arthroscopy procedure a pressure fault error message populated on the ar-6480 dualwave pump during use.
The facility followed the proper steps and switched out the pump tubing.
The pump was re-started and was working properly again with the new pump tubing.
At the end of the procedure as the surgeon was removing the scope, over-pressure and extravasation within the patient's joint was discovered.
Normal suction was used to remove all excess saline and the case was completed.
The patient was closely monitored for two hours post-op and no further action or intervention was needed as the swelling had reduced.
|