HEART VALVES SANTA ANA EVOLUT PRO PLUS DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number D-EVPROP2329US
Device Problems Device Difficult to Setup or Prepare (1487); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/07/2021
Event Type  malfunction  
Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that prior to the implant of this transcatheter bioprosthetic valve, while loading using this delivery catheter system (dcs), resistance was felt.The dcs appeared to have torn at the capsule.The dcs was not used inside the patient.The dcs was exchanged for a new one to completed the procedure.There was no patient involvement.
 
Manufacturer Narrative
Additional information was received that the valve loader was experienced, there were no difficulties noted in loading the valve.Th ere was no additional or unusual force used to load the valve.No adverse patient effects were reported.B5-updated event description.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Conclusion: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.The delivery catheter system (dcs) was returned to medtronic for analysis.The capsule of the dcs appeared intact with no evidence of a tear.As part of the analysis procedure, an 29mm evolutpro+ valve was successfully loaded onto the dcs.No procedural images were provided to medtronic for review.The reported event indicates that while loading the valve onto the dcs, resistance was felt.When the valve was loaded, a bulge was noted.A tear was suspected in-between the capsule flare and the actual braided capsule.As no procedural images were provided, the reported difficulties loading the valve could not be confirmed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information was received reporting that when the valve was loaded a bulge was noted.A tear was suspected in-between the capsule flare and the actual braided capsule, however the tear was not confirmed.No adverse patient effects were reported.B5-updated event description.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product analysis: upon receipt at medtronic¿s quality laboratory, the handle appeared intact.The device was received with the capsule partially opened.The deployment knob appeared to retract and advance the capsule.The trigger moved to fully advanced and retracted positions and locked in place when released.The tip-retrieval mechanism appeared intact.The device was returned with the end cap/screw gear snap fit connected.The capsule appeared intact with no evidence of damage.The different sections of the capsule were inspected underneath a microscope and it appeared to be intact with no evidence of damage.The inner member shaft and spindle hub appeared intact with no evidence of damage.A 29mm evolutpro+ valve was successfully loaded onto the delivery catheter system (dcs).On retraction of the capsule via the rotation of the deployment knob, the valve deployed as expected.The reported event for capsule separation and difficult to load could not be confirmed in the analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT PRO PLUS DCS
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
HEART VALVES SANTA ANA
1851 e deere ave
santa ana CA 92705
MDR Report Key11652595
MDR Text Key245106146
Report Number2025587-2021-01117
Device Sequence Number1
Product Code NPT
UDI-Device Identifier00763000207854
UDI-Public00763000207854
Combination Product (y/n)N
PMA/PMN Number
P130021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 06/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/09/2022
Device Model NumberD-EVPROP2329US
Device Catalogue NumberD-EVPROP2329US
Device Lot Number0010486973
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2021
Date Manufacturer Received06/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age75 YR
Patient Weight55
-
-