Catalog Number JC4041P |
Device Problem
Incomplete or Inadequate Connection (4037)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Initial reporter address: (b)(6).Device manufacturer address: (b)(4).The devices were not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported there was difficulty spiking an unspecified quantity of y-type tur/bladder irrigation sets into a bag.This occurred during an unspecified process step.It was further reported the spike was "shorter and less sharp" than usual.During insertion into a 3000cc baxter bag of fluid, the spike "did not seem to go in very far".There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: two (2) actual samples were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed including pressure testing and clear passage testing and the results were satisfactory.Additional priming was performed and no issues noted.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
|
|