SYNTHES GMBH SAGITTAL SAW ATTACHMENT FOR PEN DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.039 |
Device Problem
Vibration (1674)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/23/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device had component damage and vibration.Therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to maintenance, which is user error/misuse/abuse.Udi ¿ (b)(4).
|
|
Event Description
|
It was reported from the (b)(6) that during service and evaluation, it was determined that the sagittal saw attachment device had component damage, moving parts of the device were not moving smoothly, vibration, the device was worn, corrosion/rusting/pitting, and the mechanics were damaged.It was further determined that the device failed pretest for check for free movement and general condition.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred on the (b)(6) day of an unknown month in 2021.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|