It was reported that the health care professional found a red alert on this cardiac resynchronization therapy pacemaker (crt-p).
The alert detected was for a high out-of-range (ooo) left ventricular (lv) lead pacing impedances.
It was noted a couple months ago the impedances have been rising intermittently measuring as high as 1424 ohms.
Boston scientific technical services (ts) discussed having the patient complete a patient initiated interrogation and bring in for clinic assessment.
Ts recommended performing isometrics and pocket manipulation to see if noise is present or if they can get any measurements.
The clinic check will reset the alert condition.
At this time this crt-p and lv lead remains in service.
No adverse patient effects were reported.
|