The affected device was examined on site by the dräger service technician and the log file was made available for further investigation at the manufacturer¿s site.Based on the log file analysis, it could be confirmed, that the device performed several warm starts at the reported time on the 19th of march 2021 until it was switched off by the user at 12.55 p.M.A faulty o2 mixer cartridge was identified to be the root cause of the warm starts.During a warm start, the emergency breathing valve opens, allowing the patient to breathe spontaneously.A warm start is accompanied by an audible alarm and the display is turned dark.If the boot sequence of the warm start is successfully completed (duration approx.8 seconds), ventilation is continued with the previous parameters.In case of an unsuccessful warm start, the warm starts are repeated until the device is switched off.The device reacted as specified and attempted to remedy an internal deviation by warm starts until the device was switched off.The warm starts were alarmed accordingly.Replacing the o2 mixer cartridge remedied the issue; the device was finally tested according to the manufacturer¿s specification and made available for further use at the customer¿s site.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|