The affected device was examined by dräger service technician onsite and the log file and service report were provided for further investigation at the manufacturer¿s site.The log file analysis showed that the ventilation unit performed a warm start two times during ventilation on the 3rd of march 2021 caused by a temporary time-out in the software task processing.Such a warm start is a specified behavior to reset the micro processing system to a specified state.The ventilation was automatically continued with the latest settings after the warm start had been completed.As per log file the accompanying alarm message «ventilation unit restarted» was posted by the cockpit infinity c500.As a safety feature of the system, the safety software analyses and verifies proper function of the device.If the safety software detects an internal failure concerning the ventilator, a local warm start of the ventilation unit would be triggered, and the emergency-breathing valve would open to ambient allowing the patient for spontaneous breathing.Alarms would be posted in order to alert the user of the situation.The device reacted as specified to a detected temporary deviation and initiated a warm start in order to remedy the issue.The pba m16.2 and the cf card were replaced and the software was reinstalled as a precautionary measure; the device was successfully tested according to the manufacturer's specification and returned back into service.There were no patient health consequences reported.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|