Model Number PULSAR
Device Problem Output Problem (3005)
Patient Problem Failure of Implant (1924)
Event Date 04/12/2021
Event Type  Injury  
Manufacturer Narrative
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event Description
The user has ongoing issues with fluctuating impedance levels.
 
Manufacturer Narrative
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is plausible.However to determine an exact root cause a device investigation of the explanted device is necessary.Re-implantation is considered but no date has been scheduled yet.
 
Event Description
The user has ongoing issues with fluctuating impedance levels.The user reports that the sound quality changes as does the loudness throughout the day.
 
Manufacturer Narrative
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue wire breakages.The problems given in the recipient report appear to match the damage found.This is a final report.
 
Event Description
The user has ongoing issues with fluctuating impedance levels.This has been managed by the clinic for the last few years.However, the user_s hearing performance is declining and now day to day hearing is affected.Re-implantation was performed on (b)(6)2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL COMBI 40+ COCHLEAR IMPLANT SYSTEM
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key11664016
MDR Text Key245344605
Report Number9710014-2021-00264
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 11/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberPULSAR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2021
Was the Report Sent to FDA? No
Date Manufacturer Received04/12/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age55 YR
Patient SexFemale
-
-