Model Number PULSAR |
Device Problem
Output Problem (3005)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 04/12/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
The user has ongoing issues with fluctuating impedance levels.
|
|
Manufacturer Narrative
|
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is plausible.However to determine an exact root cause a device investigation of the explanted device is necessary.Re-implantation is considered but no date has been scheduled yet.
|
|
Event Description
|
The user has ongoing issues with fluctuating impedance levels.The user reports that the sound quality changes as does the loudness throughout the day.
|
|
Manufacturer Narrative
|
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue wire breakages.The problems given in the recipient report appear to match the damage found.This is a final report.
|
|
Event Description
|
The user has ongoing issues with fluctuating impedance levels.This has been managed by the clinic for the last few years.However, the user_s hearing performance is declining and now day to day hearing is affected.Re-implantation was performed on (b)(6)2021.
|
|