Catalog Number 5000-01-03
Device Problems Inaccurate Flow Rate (1249); Short Fill (1575)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/25/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the routine 6th monthly service on the arctic sun device was performed and the arctic sun device did not reach the desired parameters (42 degrees) to pass the functionality test.Per sample evaluation, error 62 was displayed during calibration and screen protector needed and water level on unit was low.
 
Manufacturer Narrative
Upon further review, bd has determined that this mdr was reported in error as it was found to be a cascading failure of an event previously reported.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the routine 6th monthly service on the arctic sun device was performed and the arctic sun device did not reach the desired parameters (42 degrees) to pass the functionality test.Per sample evaluation, error 62 was displayed during calibration and screen protector needed and water level on unit was low.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC SUN® 5000
Type of Device
ARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key11665117
MDR Text Key245384611
Report Number1018233-2021-02116
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5000-01-03
Was Device Available for Evaluation? No
Date Returned to Manufacturer03/25/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/18/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-