Updated fields: b4, e2, g4, g7, h2, h6 (type of investigation), h10.Corrected fields: d11, e3, h4, h6 (medical device - problem code).Analysis of production: (3331/213) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 24 month product complaint trend data for the period (apr 2019 through mar 2021) was reviewed.There were no triggers identified for the review period.
|