SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
Model Number 72200616 |
Device Problem
Complete Blockage (1094)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/22/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reported device, intended for use in treatment, was received for evaluation.
There was a relationship found between the returned device and the reported incident.
A visual inspection was performed on the product and no issues were observed.
A functional evaluation revealed a blade stall error.
A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
A complaint history review concluded this was a repeat issue.
The complaint was confirmed and the root cause has been associated with a mechanical component failure.
Factors that could have contributed to the reported event include corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time or one or more of the motor phases shorting out.
No containment or corrective actions are recommended at this time.
|
|
Event Description
|
It was reported that the shaver was clogging.
It is unknown if the event happened during a procedure, no delay was reported and there was backup available.
No patient injuries reported.
Results of investigation have concluded that this unit had a motor stalled which makes it a reportable event.
|
|