SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
Model Number 72200616 |
Device Problem
Complete Blockage (1094)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/22/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issues were observed.A functional evaluation revealed a blade stall error.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.The complaint was confirmed and the root cause has been associated with a mechanical component failure.Factors that could have contributed to the reported event include corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time or one or more of the motor phases shorting out.No containment or corrective actions are recommended at this time.
|
|
Event Description
|
It was reported that the shaver was clogging.It is unknown if the event happened during a procedure, no delay was reported and there was backup available.No patient injuries reported.Results of investigation have concluded that this unit had a motor stalled which makes it a reportable event.
|
|