Model Number V173 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/22/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
This device has been returned and is in the process of device analysis.
|
|
Event Description
|
It was reported that an alert was received on the home monitoring system which found this cardiac resynchronization therapy pacemaker (crt-p) to be in safety mode.It was recommended that this device be replaced.The device was subsequently explanted and replaced successfully.No additional adverse patient effects were reported.
|
|
Event Description
|
It was reported that an alert was received on the home monitoring system which found this cardiac resynchronization therapy pacemaker (crt-p) to be in safety mode.It was recommended that this device be replaced.The device was subsequently explanted and replaced successfully.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode due to system resets.It was confirmed that brady therapy remained available.The system resets occurred during a telemetry session and caused the device to enter safety mode.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption related to telemetry attempts and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.This device has been returned and is in the process of device analysis.
|
|