SMITH & NEPHEW, INC. VAS HD AC 4MM X 30 DEG; NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) Back to Search Results
Model Number 72202959
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/23/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that the distal lens of the scope was cracked.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6: the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found distal tip damage, fiber damage, and a cracked and scratched negative lens.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.The complaint was confirmed.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAS HD AC 4MM X 30 DEG
Type of Device
NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key11669476
MDR Text Key245566714
Report Number3003604053-2021-00155
Device Sequence Number1
Product Code EOB
UDI-Device Identifier00885554018783
UDI-Public00885554018783
Combination Product (y/n)N
PMA/PMN Number
K043395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 05/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72202959
Device Catalogue Number72202959
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2021
Date Manufacturer Received05/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-