U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS HANAULUX 3000; LAMP, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS HANAULUX 3000; LAMP, SURGICAL Back to Search Results
Model Number ARD567904999
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/12/2021
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.
 
Event Description
On (b)(6) 2021 getinge became aware of an issue with hanaulux 3000 surgical light.The paint was chipping from a fork.There was no injury reported, however, we decided to report the issue in abundance of caution as any paint particles falling off into sterile field or during procedure may cause contamination.
 
Manufacturer Narrative
Getinge became aware of an issue with hanaulux 3000 surgical light.The paint was chipping from a fork.There was no injury reported, however, we decided to report the issue in abundance of caution as any paint particles falling off into sterile field or during procedure may cause contamination.Based on the information collected to date it was established that when the event occurred, the surgical light did not meet the manufacturer¿s specification, since appearance of chipped paint could be considered as technical deficiency, and in this way device contributed to event.There is no information if the device was or was not being used for patient treatment when the event took place.All maquet sas products comply with: ¿ iec 60601-1 ed.2.0 & ed.3.0 general requirements for basic safety and essential performance.¿ iec 60601-2-41 particular requirements for the safety of surgical luminaires and luminaire for diagnosis.¿ paint definition pfc066.This procedure defines maquet sas¿s requirements for all painted parts.¿ disinfection products test: the aim of these tests is to detect any incompatibility with disinfectant.The paint chip or paint damages are due to: impacts, collisions (abnormal use).The operating manual includes the instructions to pre-position the arms prior to use, in order to prevent damages.To prevent any similar incident, it is recommended to avoid the collisions between devices.Visual inspections during the cleaning allow detecting the painting defect, we recommend to perform corrective maintenance to rectify the default after its detection.Minor paint chip can be repaired with touch up paint, nevertheless, the parts impacted by serious damage must be replaced.It is also recommended to avoid excessive friction during cleaning or inappropriate cleaning products.We believe that remaining devices are performing correctly in the market.Given the circumstances and the fact that there is no apparent trend in complaints of this nature we shall continue to monitor for any further events of this nature and do not propose any further action at this time.
 
Event Description
Manufacturer reference number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANAULUX 3000
Type of Device
LAMP, SURGICAL
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
MDR Report Key11669979
MDR Text Key245526386
Report Number9710055-2021-00131
Device Sequence Number1
Product Code FTD
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberARD567904999
Device Catalogue NumberARD567904999
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/26/2021
Patient Sequence Number1
-
-