• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROCLAIM 5 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROCLAIM 5 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG Back to Search Results
Model Number 3660
Device Problems Device Damaged by Another Device (2915); Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Date 12/29/2020
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.
 
Event Description
Related manufacturer reference number: 1627487-2021-13235.It was reported the patient experienced heating at the ipg and lead site during an mri procedure.The scs was placed in mri mode prior to the procedure.Further troubleshooting may be pending to address the issue.
 
Manufacturer Narrative
Correction to section h6 - the medical device problem code should have been 2915 instead of 3022.During processing of this complaint, attempts were made to obtain patient weight.Further information was requested but not received.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Additional information indicated that troubleshooting showed no abnormalities with patient's scs system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 5 ELITE IMPLANTABLE PULSE GENERATOR
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
MDR Report Key11683462
MDR Text Key245977537
Report Number3006705815-2021-01793
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067031419
UDI-Public05415067031419
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 05/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/18/2021
Device Model Number3660
Device Catalogue Number3660
Device Lot NumberA000091983
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
SCS PADDLE LEAD
Patient Outcome(s) Other;
-
-